Siemens Software version syngo CT 2005C ユーザーズマニュアル

ページ / 321
Abdomen
117
 
Hint
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for 
“repeat“, e.g., when both non-contrast and contrast 
studies are required.
AbdFast
kV
130
Effective mAs
60
Rotation time
0.8 sec
Slice collimation
5.0 mm
Slice width
8.0 mm
Feed/Rotation
20.0 mm
Pitch Factor
2.0
Increment
8.0 mm
Kernel
B41s
CTDI
Vol
6.47 mGy
Effective dose
Male: 2.59 mSv
 
Female: 3.29 mSv
Contrast medium IV injection
Start delay
25 – 30 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 117  Friday, April 8, 2005  9:55 AM