NPPV Mode Setting
SPONT/T
N/A
SPONT/T
SPONT
◆◆
NPPV Patient Type
ADULT
N/A
ADULT
PED
◆◆◆
NEO
◆◆◆
NPPV Respiratory 
Rate Setting
12
◆◆◆◆
1
1  -  80
BPM
NPPV EPAP Setting
5
◆◆◆◆◆
1
2  -  25
cmH
2
O
(hPa)
NPPV IPAP Setting
5
◆◆◆◆◆
1
2  -  35
cmH
2
O
(hPa)
NPPV Inspiratory Time 
Setting
1.0
◆◆◆
0.1
0.1 - 9.9
sec
NPPV Rise Time
0.1
◆◆◆
0.1
0.1 - 0.9
sec
NPPV I-Trigger Type
FLOW
◆◆
N/A
FLOW
◆◆
AUTO
◆◆
N/A
NPPV Flow I-Trigger 
Setting
3.0
◆◆◆
0.1
0.5 - 20.0
LPM
NPPV E-Cycle Type
%FLOW
N/A
%FLOW
AUTO
◆◆
N/A
NPPV %Flow 
Expiratory Cycle 
Setting
25
◆◆◆◆
1
10 - 80
%
NPPV Oxygen 
Concentration Setting
21
◆◆◆◆
1
21 - 100
%
NPPV Apnea Rate 
Setting
12
◆◆◆◆
1
1  -  80
BPM
NPPV Low Inspiratory 
Pressure Alarm Limit 
Setting
3
◆◆◆◆◆
1
3  -  105
cmH
2
O
(hPa)
NPPV Low EPAP 
Alarm Limit Setting
0
◆◆◆◆◆
1
0  -  25
cmH
2
O
(hPa)
NPPV Low Tidal 
Volume Alarm Limit 
Setting
0
◆◆◆◆◆
1
0 - 2500
mL
NPPV High 
Respiratory Rate 
Alarm Limit Setting
150
◆◆◆
1
0 - 150
BPM
NPPV Low Minute 
Volume Alarm Limit 
Setting
1.00
◆◆
0.01 for 
0.00 to 9.99
0.1 for 10.0 
to 60.0
0.00 - 60.0
L
NPPV Apnea Interval 
Alarm Limit Setting
20
◆◆◆◆
1
10 - 60
sec
Non-Invasive Positive Pressure Ventilation Settings Report (Continued)
Description
Example
Resolution Range
Units
Comments