Philips V200 ユーザーズマニュアル

ページ / 328
Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-19
Measured End 
Expiratory Pressure 
Data
2.0
◆◆◆
0.1
-20.0-99.9
cmH
2
O
(hPa)
The End Exp. value 
from the Patient 
Data screen.
Measured End 
Inhalation Pressure
24.0
◆◆
0.1 for -20.0 to 99.9
1 for 100 to 130
-20.0 - 130 cmH
2
O
(hPa)
The Plateau value 
from the Patient 
Data screen.
Measured Tidal 
Volume
468
◆◆◆
1
0 - 2500
mL
The Tidal Volume 
value from the 
Patient Data 
screen.
Measured 
Spontaneous 
Minute Volume
0.00
◆◆
0.01 for 0.00 to 
9.99
0.1 for 10.0 to 60.0
0.00 - 60.0 L
The Spont VE value 
from the Patient 
Data screen.
Measured Minute 
Volume
5.83
◆◆
0.01 for 0.00 to 
9.99
0.1 for 10.0 to 60.0
0.00 - 60.0 L
The Total VE value 
from the Patient 
Data screen.
Measured 
Spontaneous 
Breath Rate
0.0
◆◆◆
0.1 for 0.0 to 9.9
1 for 10 to 150
0.0 - 150
BPM
The Spont Rate 
value from the 
Patient Data 
screen.
Measured Total 
Breath Rate
12.0
◆◆
0.1 for 0.0 to 9.9
1 for 10 to 150
0 - 150
BPM
The Total Rate 
value from the 
Patient Data 
screen.
Measured Rapid 
Shallow Breathing 
Index
5
◆◆◆◆◆
1
0 - 500
BPM/L
The F/Vt value from 
the Patient Data 
screen.
Measured I:E Ratio
1:4.1
0.1 for 9.9:1 to 1:9.9
1 for 1:10 to 1:99
4.1:1-1:99
N/A
The I:E Ratio 
display from the 
Patient Data 
screen.
Measured Patient 
Leak
0.0
◆◆◆
0.1 for 0.0 to 99.9
1 for 100 to 140
0.0 - 140
LPM
The Pt Leak 
display.
Measured Percent 
of Breaths 
Triggered by the 
Patient (NPPV 
Spont/T mode only; 
otherwise 
◆◆◆◆◆◆)
56.2
◆◆
0.1 for 0.0 to 99.9
1 for 100
0.0 - 100
%
The Pt Trigger 
display on the 
NPPV Monitor 
screen.
Monitored Oxygen 
Concentration
55.1
◆◆
0,1 for 0.0 to 99.91 for 
100
0.0 - 100
%
The % O2 display 
from the Patient 
Data Screen.
Ti/Ttot
0.23
◆◆
0.01
0.00 - 1.00 N/A
The Ti/Ttot display 
from the 
Mechanics Patient 
Status Screen
Patient Data Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-5: Patient Data Report (Sheet 2 of 4)