HI 98401 measures fluoride from 0.05 mg/L to 1.9 g/l in 5 distinct
scales. With auto-ranging HI 98401 automatically selects the range
that provides the best resolution. All results are displayed directly in
fluoride (F 
-----
) concentration. In order to guarantee maximum repeat-
ability, the measurements are performed utilizing, the optional FC
301B fluoride specific electrode and the separate HI 5313 reference
electrode.
HI 98401 compensates for temperature automatically from 5 to 35°C
using the optional HI 7662 stainless steel temperature probe. The
temperature mensured together with fluoride concentrations are dis-
played on the large LCD simultaneously.
All operations are microprocessor controlled for added precision and
simplicity. The calibration is automatic at one or two points. The first
calibration point is in the middle of the range at 100 mg/L (ppm). The
slope can then be selected for low fluoride measurements at 10 mg/L,
or higher contents at 1000 mg/L.
HI 98401 comes in a rugged waterproof casing so that measurements
may be performed indoors as well as outdoors. Four common AA type
batteries provide up to 120 hours of continuous operation.
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any noticeable damage, notify your Dealer or the nearest
Hanna office immediately.
Each meter is supplied with:
• AA type battery (4 pcs)
• Rugged carrying case
• Instruction manual
Note: Save all packing materials until you are sure that the instru-
ment functions correctly. Any damaged or defective item must
be returned in its original packing materials together with the
supplied accessories.
4
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
1) Temperature probe connector
2) Reference electrode connector
3) Fluoride electrode connector (BNC)
4) Liquid Crystal Display (LCD)
5) ON/OFF button
6) CAL button
7) BUF button
8) CFM button