Hanna Instruments hi 98401 ユーザーズマニュアル

ページ / 8
5
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Scale
F
-----
0.050 to 0.500 ppm
0.50 to 5.00 ppm
5.0 to 50.0 ppm
50 to 500 ppm
0.50 to 1.90 g/L
Temp. 0.0 to 50.0 °C
Resolution
F
-----
0.001 ppm
0.01 ppm
0.1 ppm
1 ppm
0.01 g/L
Temp. 0.1 °C
Accuracy
F
-----
±5% of reading or ±0.02 ppm (whichever
greater) with less than ±3°C difference from
calibration temperature; or
±10% of reading or ±0.02 ppm (whichever
greater) with temp. difference greater than ±3°C
Temp. ±0.5 °C
EMC
F
-----
± 2% f.s.
Deviation Temp. ±0.5°C
Calibration
Automatic 1 or 2 points; first point at 100 ppm
second point at 10.0 or 1000 ppm
Temperature
Automatic 5 to 35 °C (41 to 95°F)
Compensation
with temperature probe attached
Electrode
FC 301B, fluoride with 1m (3') cable and BNC
connector
HI 5313, reference with 1m (3') cable
Temperature
HI 7662 with 1 m (3') cable
Probe
Power supply
4 x 1.5V AA alcaline batteries
120 hours of continuous use
Environment
0 to 50°C (32 to 122°F); 100% RH
Dimensions
196 x 80 x 60 mm (7.7x3.1x2.4")
Weight
450 g (15.9 oz.)
6
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDE
INITIAL PREPARATION:
The meter is supplied with 4 pieces of 1.5V AA-type alkaline batteries.
Remove the battery cover on the back of the meter and insert the
batteries while paying attention to their polarity.
Attach the fluoride electrode (#3 on
page 4), the reference electrode (#2)
and the temperature probe (#1) to their
respective connectors on top of the meter.
Press the ON/OFF key to turn the meter on.
Note: The temperature probe can also be used on its own to measure
the temperature independently.
Note: If the temperature probe is not con-
nected, the lower display will indicate
25.0°C with the “°C” symbol blinking.
MEASUREMENT OF FLUORIDE ION CONCENTRATION
Note: Please ensure the meter is calibrated prior to proceeding any
further. In order to obtain accurate results, calibrate the meter
at a temperature no more than ±3°C away from the sample
to be measured subsequently.
Note: If the fluoride electrode is new, or it has not been used for a
few days, it must be soaked in a solution of HI 70702 (10 ppm
of F
-----
) for at least 3 hours prior to measurement.
) for at least 3 hours prior to measurement.
) for at least 3 hours prior to measurement.
) for at least 3 hours prior to measurement.
) for at least 3 hours prior to measurement.
• Pour about 10-20 mL of the sample in a clean beaker and add the
same quantity of HI 7023 (TISAB) solution to the beaker. Mix the
solution for a few minutes.
• Rinse the electrodes with deionized water and dry them attentively
with a non-abrasive cloth.
• Immerse the fluoride and the reference electrode as well as the
temperature probe to a depth of at least 1 cm
(½”) in the prepared sample. Make sure that
the electrodes are not touching the bottom of
the beaker. Position the temperature probe close
to the tip of the other electrodes.
Temperature
probe
F Electrod Ref.