6.2 Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic discharge
(ESD)
IEC 61000-4-2
±2.0 kV, ±4.0 kV, ±6.0kV,
±8.0 kV contact discharge
±2.0 kV, ±4.0 kV, ±8.0 kV,
±15.0 kV air discharge
±2.0 kV, ±4.0 kV,
±6.0 kV, ±8.0 kV
contact(56% RH)
±2.0 kV, ±4.0 kV,
±8.0 kV, ±15.0 kV
air (56% RH)
RF
electromagnetic
field immunity test
IEC 61000-4-3
10 V/m
10 V/m
Power
frequency
magnetic fields
IEC 61000-4-8
30 A/m
30 A/m
Warning:
1. Medtrum MD-SY-011C system is not designed to be used in an environment
with high voltage, high-intensity magnetic field, where the intensity of EM
DISTURBANCES is high.
2. Portable RF Communications equipment should be used no closer than 30
cm (12 inches) to any part of the Medtrum products. Otherwise, degradation