Lecroy WaveJet 354A 500 MHz 4 Channel Oscilloscope WaveJet 354A データシート

製品コード
WaveJet 354A
ページ / 60
WaveJet 300A Series 
16
 
917976 Rev A
 
4. LIMITATIONS ON REMOTE FUNCTION 
The following limitations are imposed when Remote Control is selected. 
•  Delimiters permit the following settings. 
Table 4-1. Delimiters 
State 
Delimiter (GPIB and TCP/IP) Delimiter 
(USB) 
Recv EOI 
CR 
Send LF+EOI  LF+CR 
 
•  Panel Lock mode is entered. 
•  The setting is changed to the following once Remote Control mode is entered. 
Table 4-2. Setup in Remote Control mode 
Item Setup 
Panel 
Help display 
Off 
Offset Setting 
Division 
Utilities/Config. 1/2 
Power Off 
Never 
Utilities/Config. 1/2/Power 
Management 
Panel Lock 
Lock 
Utilities/Config. 2/2 
Volts/div Coarse  Channel 
Menu 
2/2 
Zoom Off 
ZOOM 
Line Number (TV Standard = NTSC) 
263/262 
Trigger Menu/Type = TV/TV Setting 
Line Number (TV Standard = PAL) 
625 
Trigger Menu/Type = TV/TV Setting