6. REMOTE COMMANDS 
This list of remote commands is alphabetized by short form. 
Horizontal ACQ 
  
Command/Query 
DESCRIPTION 
The ACQ command selects the acquisition mode. The ACQ? 
query returns the current selection of the acquisition 
mode. 
COMMAND SYNTAX 
ACQ <mode> 
<mode>:={NORMAL, PEAK, AVERAGE} (When XYDS is YT) 
<mode>:={NORMAL, AVERAGE} (When XYDS is XYTRG) 
<mode>:={NORMAL} (When XYDS is XY) 
QUERY SYNTAX 
ACQ? 
RESPONSE FORMAT 
<mode> 
REMARKS 
ACQ AVERAGE sets ROLL OFF automatically. 
 
 
Acquisition ASET 
  
Command 
DESCRIPTION 
The ASET command provides the auto setup. 
COMMAND SYNTAX 
ASET 
QUERY SYNTAX 
  
RESPONSE FORMAT 
  
 
 
Horizontal AVGCNT 
  
Command/Query 
DESCRIPTION The 
AVGCNT command sets the number of events to average. 
AVGCNT? query returns the number set. 
COMMAND SYNTAX 
AVGCNT <number> 
<number>:={2, 4, 8, 16, 32, 64, 128, 256} 
QUERY SYNTAX 
AVGCNT? 
RESPONSE FORMAT 
<number> 
REMARKS 
AVGCNT command reports device-specific error (DDE) and 
is not performed when ACQ is not AVERAGE. 
 
 
Vertical BWL 
 Command/Query 
DESCRIPTION The 
BWL command enables or disables the bandwidth limit 
(low-pass filter) of the specified input channel. The 
BWL? query returns the selected status. 
COMMAND SYNTAX 
<channel>: BWL <mode> 
<channel>: = {C1, C2, C3, C4} (WJ354/334/324/314) 
             : = {C1, C2} (WJ352/332/322/312) 
<mode>: = {OFF, 100M, 20M} (WJ354/334/352/332) 
          : = {OFF, 20M} (WJ324/314/322/312) 
QUERY SYNTAX 
<channel>:BWL? 
RESPONSE FORMAT 
<mode> 
The current bandwidth limit status (low-pass filter)