4.1. 
Interior of Test Chamber 
  Remove the sample pan support from its separate box and packing foam.   
  Save the box and foam so that the support can be packaged when the instrument 
is returned to the factory for annual calibration. 
 
NOTE: Shipment of the instrument with the pan support in place will 
damage the ceramic oven and the balance.  Damage caused by 
improper shipment is not covered by the warranty. 
 
  Open the lid of the MAX 5000XL by lifting the 
handle just above the display. 
  Install the pan support gently but firmly into the 
instrument.  The stem of the pan support fits 
snugly into the balance. 
  Check the clearance between the pan support 
arms and the heat shield by observing the 
space between the arms and the heat shield.  
The clearance should be between 1.5mm to 
4.75mm, [1/16” to 3/16”] and equal for all five 
arms of the pan support.  If the clearance is 
considerably different for one arm or the pan 
support appears to be crooked, call AZI 
Customer Service at 800-528-7411 or 
602-470-1414. 
 
CAUTION - Avoid excess pressure from the top or sides of the pan 
support that can damage the force balance.  Only install or remove 
the pan support when the instrument is powered OFF.