4.2. 
AC Power Connection  
 
 
 
 
 
The MAX 5000XL can be used with a 100-130 VAC or a 220-240 VAC input power 
source.  Check the setting on the fuse holder, located at the rear of the instrument, and 
make sure it is set for the correct voltage for your environment.  
 
Ensure the AC power receptacle is set for the correct voltage for 
your environment.  For best results, AZI recommends that the 
instrument is provided with its own DEDICATED electrical power 
outlet.  Do not put it on a circuit with motors, blenders, heaters, 
coolers, grinders, or other high current electrical devices. 
 
 
To ensure the entry module is set for the correct voltage:   
  Locate the fuse holder on the rear of the instrument. 
  Pull the fuse access cover open and down. 
  Pry the fuse holder from the power entry 
module. 
  Ensure only 8 amp timed fuses are installed 
for 100-130 VAC operation or only 4 amp 
timed fuses are installed for 220-240 VAC 
operation. 
  Reinstall the fuse holder into the power entry module with the correct voltage 
showing, as in the picture above which shows 115V configuration. 
  Ensure the power switch, located next to the fuse holder, is in the OFF position. 
  Remove the power cord from the packing material and insert the rectangular end 
into the power receptacle, located on the other side of the fuse holder. 
 
The power source outlet used by the MAX 5000XL MUST BE 
GROUNDED
.  (If you are not sure if the socket is grounded, check 
with your plant electrician before proceeding).  The MAX 5000XL 
will appear to function correctly even if the socket is not 
grounded, but there is a danger of possible electrocution!  If 
grounded outlets are not available, consider using a ground 
fault interrupter to protect personnel against electrical shock.
 
Power Requirements:  100-130 VAC, 50/60 Hz @ 8 Amps 
– fuse is F8A T250V 
220-240 VAC, 50/60 Hz @ 4 Amps 
– fuse is F4A T250V 
Fused on/off switch is part of an EMI power entry module