  After power-up, the balance temperature will be stabilized for high temperature 
testing, which may take up to 2 hours.  Wait for the balance temperature stability 
icon on the main screen to change from unstable   to stable   before performing 
the balance calibration or low moisture/solids/ash testing (<1%).  See section 6.1 
Balance Temperature Stability 
on page 64 for more information. 
 
4.3. 
Contrast Adjustment 
If necessary, adjust the display 
contrast controls for better viewing of 
the display.  The left most soft key 
below the display has the word 
CONTRAST above it.  Press this key 
and the display changes to instructions 
on setting contrast.  Use the up and 
down arrows on the key pad until the 
display is acceptable and press the 
[ESC] key to save the setting and exit 
back to the Main Test Screen. 
The LCD will darken after a few 
minutes of inactivity but will brighten 
with any key press. 
 
4.4. 
Connecting a Printer 
An optional graphic capable printer is available from AZI (pictured 
is printer example only).  It connects to the parallel printer port at 
the rear of the instrument.  The content and time of printing for 
each report can be custom configured from the menu system.  
These same reports can be directed to a computer through the 
RS-232 serial port at the back of the instrument. 
  Unpack the printer and cable.  Load the paper according to 
the instructions in the printer manual.  Plug the large end 
cable connector into its mating socket on the 
printer, and the small end into its mating socket on 
the back of the MAX 5000XL.  Plug the power cord 
into the printer, and into a GROUNDED power 
receptacle. 
  Turn the printer power on.  (On some older 
printers, it is also necessary to place the printer 
“On Line.”) 
 
 
 
NOTE: The USB and AUX ports are not active at present.  They will 
be implemented in later revisions to the instrument