Arizona MAX-5000XL ユーザーズマニュアル

ページ / 92
AZI Customer Service 800-528-7411 or 602-470-1414  
Page 20 of 92 
4.7. 
Calibrate the Balance - 50 gram weight 
NOTE: The MAX 5000XL must be powered on until the balance temperature 
stability icon in the lower left corner of the main screen changes from unstable   
to stable   to allow the balance to stabilize before beginning the balance 
calibration.  (See section 6.1 Balance Temperature Stability 
on page 64 for more 
information).  The unit should not be moved, jarred, or otherwise disturbed in any 
way during the calibration period.  A vibration-free table will aid the calibration 
process.  If vibration is excessive, the instrument will stop the calibration process 
and indicate the reason. 
 
The balance should be calibrated: 
  Immediately after initial setup and 
warm-up, 
  Whenever the instrument is 
moved, 
  Whenever a problem is suspected, 
or 
  As your quality policy dictates. 
  From the Main Test Screen, press 
[Menu]. 
  Select the CALIBRATION 
MENU.  
  Select BALANCE 
CALIBRATION. 
  Select CALIBRATE BALANCE.  
 
The first screen reminds the operator that the instrument must be free of vibration 
during the calibration and that a 50-gram calibration weight must be available. 
  Press [GO]. 
 
If the chamber temperature is too hot, 
the screen will prompt the operator to 
“OPEN LID CHAMBER TEMPERATURE IS 
TOO HOT WAIT FOR COO
L DOWN”
 
If the chamber temperature is cool 
enough to begin the calibration, one of 
two prompts will appear:  
  If the lid is closed, the 
“OPEN LID”  
“PLACE CLEAN PAN ON PAN 
SUPPORT”
 
prompt will appear. 
  If the lid is open, the 
“PLACE CLEAN 
PAN ON PAN 
SUPPORT CLOSE LID”
 
prompt will appear.