4.8. 
Weight Test 
To ensure proper calibration and become familiar with the instrument, run a weight test.  
This test simulates moisture loss by the removal of one of two known weights.  Both a 
three and a five-gram weight are required.
 
 
  Ensure that a clean, dry, empty 
sample pan is placed on the pan 
support. 
  From the Main Test Screen, press 
[MEMORY]
  If the highlight bar is not already 
there, highlight (Factory) WEIGHT 
and press [START]
  One of several prompts may appear 
depending on the last measured 
weight.  Follow the instructions 
given until the 
“OPEN LID” “ADD 
SAMPLE…”
 prompt appears on the 
screen.  Open the lid and carefully 
place both the 5-gram and the 3-gram weights near the center of the sample pan. 
  Close the lid. 
  After the Main Test Screen is displayed, gently open the lid, carefully remove the 
3-gram weight, and gently close the lid. 
  Wait for the end of the test. 
  If this is the first test of the day, 
discard the results of this test for 
stability reasons and perform the 
test again. 
  Record the final test result.  The 
value should fall between 
37.4800% and 37.5200%. 
  If the results are out of tolerance, 
repeat the balance calibration and 
weight test.  If the result is still out of 
tolerance, call AZI customer service 
at 800-528-7411 or 602-470-1414 
for assistance.