Arizona MAX-5000XL ユーザーズマニュアル

ページ / 92
AZI Customer Service 800-528-7411 or 602-470-1414  
Page 40 of 92 
 
5.3. 
Calibration Menu 
 
Balance Calibration 
Balance calibration is explained and performed as part of instrument setup, see page 
20.  The sequence is repeated here for continuity. 
The balance should be calibrated: 
  If the instrument has been turned off for several hours or in storage, 
  After the instrument is moved, 
  Whenever a problem is suspected, and/or 
  As often as required by your quality policy. 
 
NOTE:  The MAX 5000XL must be powered on until the balance 
temperature stability icon in the lower left corner of the main screen 
changes from unstable   to stable   to allow the balance to 
stabilize before beginning the balance calibration.  (See section 6.1 
Balance Temperature Stability 
on page 64 for more information).  
The unit should not be moved, jarred, or otherwise disturbed in any 
way during the calibration period.  A vibration-free table will aid the 
calibration process.  If vibration is excessive, the instrument will 
stop the calibration process and indicate the reason. 
 
Calibrate Balance 
  From the Main Test Screen, press [Menu]. 
  Select the CALIBRATION MENU.  
  Select BALANCE CALIBRATION. 
  Select CALIBRATE BALANCE. 
  Press [GO].  Follow the prompts to complete the calibration. 
  The first prompt is 
“OPEN LID”
 and 
“PLACE CLEAN PAN ON PAN SUPPORT”
 
  After placing a clean pan on the pan support and closing the LID, the screen will 
display 
“CALIBRATING, PLEASE WAIT.”
 
  As soon as the tare is finished, the next prompt, 
“OPEN LID”
 and 
“LOAD 50G WEIGHT 
AND WAIT,”
 will appear. 
  As the lid is opened, the screen prompts: 
“LOAD 50G WEIGHT AND WAIT.” 
 
 
Avoid excess pressure on the scale.  Use the tweezers included 
with the MAX 5000XL to gently place the weight on the pan.