  As soon as the instrument detects weight on the balance, the prompt
 
“CLOSE LID” 
appears. 
  With the lid closed the prompt 
“CALIBRATING, PLEASE WAIT
” appears. 
  At the conclusion of the calibration, the 
“OPEN LID” “REMOVE WEIGHT”
 prompt 
appears with a PASS/FAIL indication of the calibration success. 
 
If the result is “PASS”, open the lid; remove the weight and press [PROCEED] to 
continue. 
 
If the result is “FAIL”, the calibration was unsuccessful.  Rerun the calibration. 
  If the calibration fails again, contact AZI Customer Service at 800-528-7411 or 
602-470-1414.  
  Press the ESC key three (3) times to return to the Main Test Manu. 
 
View Calibration Report 
When VIEW CALIBRATION REPORT is selected, the last balance calibration date will 
be displayed. 
 
Calibration Reminder Setup 
A prompt is available to remind you when scheduled calibrations are due.  The default 
factory settings are:   
ENABLE - No 
TIME 
– 00:00 
INTERVAL - 0 Weeks 
To turn the prompt on, off, or adjust the interval; go to the Main Test Screen and select in 
order:  MENU, CALIBRATION MENU, BALANCE CALIBRATION, and CALIBRATION 
SETUP REMINDER 
 
  Enable 
– Toggle to YES to 
enable the prompt or NO to 
prevent the prompt from 
appearing. 
  Time 
– The time the prompt is 
to appear, usually the start of 
the shift (using a 24 hour 
format). 
  Interval 
– The interval 
between calibrations (in 
weeks).  If the interval is set to 
“00”, the reminder will occur 
daily.