Arizona MAX-5000XL ユーザーズマニュアル

ページ / 92
AZI Customer Service 800-528-7411 or 602-470-1414  
Page 47 of 92 
 
Temperature Calibration Instructions 
  Ensure the instrument is turned ON 
  A Temperature Calibration Kit, [AZI P/N 
Y990-0193 for 110V or Y990-0224 for 
220V], with 9 VDC, 100 mA, 3.5 mm 
male plug positive tip adaptor, must be 
used to perform the calibration. 
  A graphics capable printer may be 
connected to the MAX 5000XL if a 
printed report is required.  
 
 
  Gently remove the pan support and ceramic balance cover. 
  Install the temperature calibration fixture so the probe 
mount and wires are as shown and gently close the 
lid. 
  Plug the disk interface cable into the matching 
connector on the TCI module box.  Plug the serial 
interface cable into the matching connector on the 
TCI module box and the serial port on the back of the 
unit.  Plug the 9VDC adaptor into the TCI module box 
and an appropriate AC outlet. 
 
 
  From the Main Test Screen select in order; MENU, CALIBRATION MENU, 
TEMPERATURE CALIBRATION 
  Highlight CALIBRATION MODULE SETUP and press [ENT].  Enter the 
appropriate data for the TCI module serial number, probe serial number and 
calibration/certification date. 
  The interface serial number and certification date are located on the TCI box 
label.