10.  ROUTINE MAINTENANCE 
 
 
The MAX 5000XL requires routine cleaning to keep the test chamber free of excess 
sample material.  Failure to keep the instrument interior clean reduces its accuracy 
and repeatability.  Sample material collecting in the test chamber prevents free 
movement of the pan support.  In addition, sample material collecting on the rim of 
the heater shroud prevents the lid from closing completely, causing erratic results. 
Material may fall into the system's interior, causing balance problems.  Routine 
cleaning of the test chamber reduces the likelihood of problems and repairs, 
extending the life of the instrument. 
 
 
10.1. 
Chamber Interior  
 
The chamber interior should be cleaned when needed to eliminate the buildup of sample 
material.  
To clean debris from the chamber’s 
interior: 
 
  Switch the main power switch OFF. 
  Use a vacuum line, portable vacuum 
cleaner, or a small paintbrush to remove 
debris from the area around the balance 
cover and pan support stem. 
 
 
 
 
 
 
 
A compressed air line or canned air is not recommended since 
debris migrates into the system's interior and settles in the 
electronic components and balance stem. 
 
Do not clean the shroud or heater element with a brush, vacuum, 
cleaner liquid, or any object.  Use the “Oven Chamber Cleaning” 
procedure.