Teledyne hfm-300 사용자 설명서

다운로드
페이지 31
Manual: 151-082010 300-302 Series
  
Page 10 of 31 
 
Span Error vs Pressure
(0.026" Sensor Tube)
y = 9.8877E-11x
3
3.4154E-07x - 
2
8.3288E-05x + 
-20.0%
-18.0%
-16.0%
-14.0%
-12.0%
-10.0%
-8.0%
-6.0%
-4.0%
-2.0%
0.0%
2.0%
0
100
200
300
400
500
600
700
800
900
1000
Pressure(psig)
Span E
rror (
%
 reading)
Mean error
max
min
2.6.2. Zero 
Check 
Turn the power supply on if not already energized.  Allow for a 1 hour 
warm-up.  Stop all flow through the instrument and wait 2 minutes.  
Caution:  Do not assume that all metering valves completely shut off 
the flow.  Even a slight leakage will cause an indication on the meter 
and an apparent zero shift.  For the standard 0-5 VDC output, adjust 
the zero potentiometer located on the inlet side of the flow meter 
until the meter indicates zero (Fig 2.3).  For the optional 4-20 mA 
output, adjust the zero potentiometer so that the meter indicates 
slightly more than 4 mA, i.e. 4.03 to 4.05 mA.  This slight positive 
adjustment ensures that the 4-20 mA current loop transmitter is not in 
the cut-off region.  The error induced by this adjustment is 
approximately 0.3% of full scale.  This zero should be checked 
periodically during normal operation.  Zero adjustment is required if 
there is a change in ambient temperature, or vertical orientation of 
the flow meter /controller. 
 
2.6.3.  High Pressure Operation 
When operating at high pressure, the increased density of gas will cause natural convection to flow 
through the sensor tube if the instrument is not mounted in a level position.  This natural convection 
flow will be proportional to the system pressure.  This will be seen as a shift in the zero flow output 
that is directly proportional to the system pressure. 
 
 
Fig. 2.4 
Fig. 2.3