Manual: 151-082010 300-302 Series
  
Page 9 of 31 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Figures 2.1/2.2, and Tables 2.1/2.2, show the 300/302 pin out. 
Pin #
Pin #
1
Signal Common
1
Do not use
2
Do not use
2
Do not use
3
Do not use
3
Do not use
4
+15 VDC
4
Do not use
5
5
Signal Common
6
Output 0-5 VDC (4-20mA)
6
Output 0-5 VDC (4-20mA)
7
Signal Common
7
Case Ground
8
Case Ground
8
Valve Override
9
Valve Override
9
-15VDC
10
10
Do not use
11
-15VDC
11
+15VDC
12
External Input
12
Signal Common
13
Signal Common
13
External Input
14
Signal Common
14
Set Point 0-5 VDC (4-20mA)
15
Set Point 0-5 VDC (4-20mA)
15
Do not use
Table 2.1
"U" Pin-Out
Table 2.2
"H" Pin-Out
 
2.6. Operation 
The standard instrument output is a 0 - 5 VDC out and the signal is proportional to the flow i.e.,    0 
volts = zero flow and 5 volts = 100% of rated flow.  The 4 - 20 mA option is also proportional to flow, 4 
mA = zero flow and 20 mA = 100% of rated flow. 
2.6.1. Operating 
Conditions 
For proper operation, the combination of ambient temperature and gas temperature must be such that 
the flow meter temperature remains between 0 and 60°C.  (Most accurate measurement of flow will 
be obtained if the flow meter is zeroed at operating temperature as temperature shifts result in some 
zero offset.)  The Hastings 300 series instrument is intended for use in non-condensing environments 
only.  Condensate or any other liquids which enter the flow meter may destroy its electronic 
components. 