Troubleshooting Guide
No fluid flow
If valve operating air pressure is too low,
the valve will not open.  Increase air
pressure to 70 psi (4.8 bar) minimum.
The reservoir air pressure may not be
high enough. Increase pressure.
The needle stroke adjustment may be
closed. Open stroke adjustment.
Material may have clogged the fluid
body or output tip adapter. Clean the
valve.
Dispensing tip retaining nut may not
have been tightened enough to unseat
the needle. Tighten the nut.
Steady drip
Remove the tip adapter/seat assem-
bly. Clean and inspect the needle.
Replace dispensing tip.
Make sure tip size is not larger than
22 gauge.
Fluid leaks out the drain hole
Fluid leaking out the drain hole on the
side of the valve indicates the needle
packings are worn. Replace needle
packings.
Inconsistent deposits
Inconsistent deposits can result if the air
pressure controlling the valve and/or
supplying the reservoir is fluctuating or if
the valve operating pressure is less than
70 psi (4.8 bar). Check to be sure air
pressures are constant and the valve
operating pressure is 70 psi (4.8 bar).
The time the valve is open must be con-
stant. Check to be sure the valve con-
troller is providing a consistent output.
Calibration Feature
The stroke control reference ring of each
741MD-SS valve is factory calibrated to
zero position. Slight internal variations in
dispensing tips may require the stroke
control to be recalibrated when tips are
changed. To do so:
1. Make a note of the current stroke
setting number.
2. Turn the calibration knob (inner)
counterclockwise one full turn.
3. Install the new dispensing tip and
ensure that the retaining nut is tight-
ened fully.
4. Turn the stroke adjustment knob
(outer) clockwise until it stops at the
zero position. 
5. Turn the calibration knob clockwise
until it stops. The stroke adjustment
is now calibrated to zero.
6. Reset stroke to the required position
noted in step 1.
In the event that the stroke reference ring
must be repositioned or reset to the zero
mark, use the following procedure:
A. Remove the dispensing tip, if installed.
B. Turn the small knob counterclock-
wise one full turn. 
C. Turn the large knob clockwise until it
stops.
If the zero on the reference ring does not
line-up with the index mark, continue with
the following steps: 
D. Loosen the small set-screw located
on the reference ring.
E. Rotate reference ring until it aligns
with the reference line on the air
cylinder body.
F.
Tighten the small set-screw to lock
the reference ring into position.
G. Install a dispensing tip and follow 
calibration procedure steps 3 to 5.