Teledyne 201A Manual Do Utilizador

Página de 31
Teledyne API Model 201A NH
3
 Analyzer Instruction Manual, 04689, Rev. B1 
2.1.1 Zero/Span Calibration with Z/S Valves or ECO Option  
The zero gas enters through the zero gas inlet port on the rear panel.  Make sure that 
the umbilical cord that makes the control connections between the analyzer and the 
converter is connected (see Figure 2 below). The zero gas valve can be activated by a 
variety of methods discussed in Section 7 of the M200A manual.  The example below 
will use the front panel keyboard to activate the valves. 
Table 2.4 - Zero Calibration Procedure - Zero Gas from rear panel zero air inlet 
 
Step No. 
 Action 
  
Comment 
1. Press 
CALZ 
The M201A enters the calibrate mode from 
sample mode.  The zero gas is supplied 
through the zero air inlet port on the rear panel.
2. Wait 
20 
min 
Wait for reading to stabilize at zero.  If the 
instrument has recently been exposed to high 
levels of NH3, you may need to wait up to 2 
hours for the instrument to return to zero.  
3. Press 
ZERO 
If you change your mind after pressing ZERO, 
you can still press EXIT here  without zeroing 
the instrument. 
4. Press 
ENTR 
Pressing ENTR actually changes the 
calculation equations. 
5. Press 
EXIT 
M201A returns to sampling.  This switches the 
ZS valves so that sample gas is coming in the 
sample port on the rear of the instrument.  
Immediately after calibration, data is not added 
to the DAS averages. 
 
Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the 
expected zero value.   
The expected value of the span gas is entered into the instrument by the following 
procedure. 
13