Página de 31
Teledyne API Model 201A NH
3
 Analyzer Instruction Manual, 04689, Rev. B1 
Table 2.6 - CALS - Span Cal Procedure - Gas from rear panel span gas inlet  
 
Step No. 
Action Comment 
1. Press 
CALS 
The M201A enters the calibrate mode.  The 
Z/S valves are activated to route span gas into 
the analyzer. 
2. Press 
 
CAL-CONV-SET 
Set the converter efficiency parameter to 1.0000, then 
press ENTR-EXIT. 
2. Wait 
20 
min 
Wait for the reading to stabilize at span value. 
3. Press 
SPAN 
If you change your mind after pressing SPAN, 
you can still press EXIT here without spanning 
the instrument. 
4. Press 
ENTR 
Pressing ENTR actually changes the 
calculation equations and causes the 
instrument to read the TN and TNx span 
concentrations. 
5. 
Press EXIT
M201A returns to sampling.  Immediately after 
calibration, data is not added to the DAS 
averages. 
 
2.1.2 TN/TNx Channel Calibration Quality 
After the TN and TNx channel calibration it is important to check the slopes and offsets 
to make sure they fall within the acceptable range.  The tolerances are different than a 
typical NOx analyzer primarily because of the performance of the M501NH ammonia 
converter. 
Slopes can vary from 2.0 to 0.7, and should be within 20% of each other with the TNx 
slope typically smaller than the TN slope. 
Offsets should be zero 
±250 mV and be within 100mV of each other. 
If the slope and offsets are within specifications, proceed on the ammonia channel 
calibration. 
 
 
15