Honeywell SC1000 Manual Do Utilizador

Página de 174
 
SC Series Instruction Manual     page 101
Hig
h-L
ev
el
 In
pu
t C
han
nel
12
12.2.4 “3-wire Voltage” 
Amplifiers with Single-wire 
Shunt Cal
Use the following wiring diagram when connecting an amplified transducer with a
3-wire voltage amplifier with single-wire shunt calibration to a High-Level Input
channel.  An example of a such device includes
transducers with Option 2d or Option 2g internal amplifiers (with shunt cal)
The High-Level Input channel’s Configuration Jumpers must be set as follows for
proper operation. See “Excitation and Signal Jumpers” on page 106.
(+)Excitation supply: “+28 VDC”
(-)Excitation supply: “GND”
Signal type: “voltage”
Signal reference: “differential”
Figure 12-3: “3-wire Voltage Amp w/Single-Wire Shunt Cal” Connection to High-Level Input 
Channel