12.2.4 “3-wire Voltage”
Amplifiers with Single-wire
Shunt Cal
Use the following wiring diagram when connecting an amplified transducer with a
3-wire voltage amplifier with single-wire shunt calibration to a High-Level Input
channel. An example of a such device includes
3-wire voltage amplifier with single-wire shunt calibration to a High-Level Input
channel. An example of a such device includes
•
transducers with Option 2d or Option 2g internal amplifiers (with shunt cal)
The High-Level Input channel’s Configuration Jumpers must be set as follows for
proper operation. See “Excitation and Signal Jumpers” on page 106.
proper operation. See “Excitation and Signal Jumpers” on page 106.
•
(+)Excitation supply: “+28 VDC”
•
(-)Excitation supply: “GND”
•
Signal type: “voltage”
•
Signal reference: “differential”
Figure 12-3: “3-wire Voltage Amp w/Single-Wire Shunt Cal” Connection to High-Level Input