eutech-instruments cyberscan con 11 conductivitytds meter Manual Do Utilizador

Página de 80
Instruction Manual 
 
CON 11/110 
 
42
8.3 
P2.0: Viewing calibration data 
This mode lets you recall the previous calibration data on each of the five measuring 
ranges. This is a “view only” mode. 
From measurement mode: 
1.  Press the SETUP key to enter Set Up menu.  
2.  Press the MI/▲ or MR/▼ key until you view 
parameter P2.0. 
3. Press 
the 
HOLD/ENTER 
key repeatedly to view the 
previous calibration data on each of the five 
measuring ranges. 
4.  Once you have scrolled through all calibration data, 
you will automatically return to the subgroup menu. 
5.  Press CAL/MEAS key to return to the measurement 
mode. 
 
NOTES: If there is no previous calibration data at a particular 
point, the primary display will show “----“.  
 
 
SETUP
cal
84.0
 p 2.0
SETUP
---
 p 2.1
SETUP
 p 2.2
HOLD
ENTER
HOLD
ENTER
µS
Figure 35: P2.0 –Viewing 
previous calibration data 
for each measurement 
range