Instruction Manual 
 
CON 11/110 
 
44
8.5 
P4.0: Auto Off  
In Program P4.0 you can activate the Auto Off option to automatically switch off the 
instrument 20 minutes after the last key selection. This feature is useful for 
conserving battery power. Under default conditions, the 
instrument has this option activated. 
1.  Press the SETUP key to enter Set Up mode.  
2.  Press the MI/▲ or MR/▼ keys until you view 
parameter P4.0.   
3.  Press the HOLD/ENTER key. The upper display 
shows “AtO” and the lower display shows “YES” or 
“NO”. 
4. Press 
MI/▲ or MR/▼ keys to select options 
• 
YES – activates the Auto Off function 
• 
NO - deactivates the Auto Off function 
5.  Press the HOLD/ENTER key to confirm and return 
to subgroup menu.  
6.  Press the CAL/MEAS key to return to measurement 
mode.  
 
 
 
 
NOTE:  
For datalogging using interval measurements, it is often desirable to use 
an AC adapter and disable the auto-off feature. 