5.2 
Manual Temperature Compensation
IMPORTANT: For manual compensation, you must disconnect the temperature probe. The
ATC annunciator will disappear from the LCD.
1. 
Press the MODE key to select TEMP mode.
2.  Press the CAL/MEAS key to enter
Temperature calibration mode. The CAL
indicator will appear above the primary
display.
3.  The primary display shows the current
conductivity/TDS reading and the
secondary display shows the default
temperature value 25 
°
C.
4.  Check the temperature of your sample
using an accurate thermometer.
5. Press 
the 
!
 or 
"
, or MI/
!
 or MR/
"
 key to
set the temperature to the measured value
obtained in step 4.  To confirm, press
ENTER/RANGE key.
6. 
Press the CAL/MEAS key to exit and return
to conductivity measurement mode.
NOTE: To exit this program without confirming
the manual temperature compensation value, DO
NOT press ENTER/RANGE in step 5. Press
CAL/MEAS instead.
Figure 17: Setting the Temperature
Compensation manually