2- 5
Getting Started and Using Masks
To calibrate the oscilloscope
To calibrate the oscilloscope
Your Infiniium should be at optimum performance when performing mask 
testing.  If your Infiniium has not been calibrated in the last year or if the 
Calibration 
∆Temp is greater than ±5 °C, you should recalibrate the oscilloscope 
before performing a mask test.  You can find the Calibration 
∆Temp in the 
Calibration dialog box by clicking on the 
Utilities menu and selecting Calibration.
1
Make sure Infiniium has warmed up for at least 30 minutes.
2
From the oscilloscope main menu click 
Utilities
, then 
Calibration
.
3
Make sure that the 
Cal Memory Protect
 check box is unchecked.
4
Click 
Start
.
Infiniium will display additional instructions and guide you through the step-
by-step calibration procedure.
Cal Memory Protect
This option protects the oscilloscope’s calibration factors from 
accidentally being changed.
Details
When checked, this option displays a list of anything that has failed 
during the calibration.  You can use the failure information to help repair 
the oscilloscope.