Getting Started and Using Masks
To choose a proper probe for Mask Testing
To choose a proper probe for Mask Testing
You can use any oscilloscope probe that fits your application.  Depending on the 
type of waveform you are testing, you may need to use one of the probe adapters.
Calibrating the probe adapter
Depending on the requirements of your Telecom or Datacom industrial 
standard, you may need to use the E2622A 75
Ω probe adapter or the E2621A 
differential probe adaptor.  Before performing a mask test, you should calibrate 
the probe adapter using the following instructions:
1
In the Channel Setup dialog box, click the 
Probes...
 button.
2
In the Probes Setup dialog box, click the 
Calibration...
 button.
3
Follow the step-by-step instructions to calibrate the probe.