Reference, Installation, and Operations Manual
Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
Initial Ethernet communications connection
103
5.3.2
Initial Ethernet communications connection
For Ethernet communications, configure these data points (listed in the table below):
5.3.3
Ethernet communications connection
Table 5-3 Data Points for Ethernet Port Configuration
Daniel MeterLink
Display Name
Display Name
Data Points, Options and Guidelines
IP Address
Data points affected:
•
Eth1IPAddr
Options:
•
Enter a value (stored as a string)
Subnet mask
Data points affected:
•
Eth1SubnetMask
Options:
•
Enter a value (stored as a string)
Default gateway
Data points affected:
•
EthDfltGatewayAddr
Options:
•
Enter a value (stored as a string)
for further
information on switch settings.
for communication troubleshooting information.
It is strongly recommended that the meter be configured using an independent (off-network)
single host computer. After configuration of the meter, the DHCP option should be disabled.
single host computer. After configuration of the meter, the DHCP option should be disabled.