Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
Initial Ethernet communications connection
 103
5.3.2
Initial Ethernet communications connection
For Ethernet communications, configure these data points (listed in the table below): 
5.3.3
Ethernet communications connection     
Table 5-3  Data Points for Ethernet Port Configuration
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
IP Address
Data points affected:
Eth1IPAddr
Options:
Enter a value (stored as a string)
Subnet mask
Data points affected:
Eth1SubnetMask
Options:
Enter a value (stored as a string)
Default gateway
Data points affected:
EthDfltGatewayAddr
Options:
Enter a value (stored as a string)
These instructions involve setting Mark III switches. Refer to 
 for further 
information on switch settings.
Refer to 
 for communication troubleshooting information.
It is strongly recommended that the meter be configured using an independent (off-network) 
single host computer. After configuration of the meter, the DHCP option should be disabled.