Instruction Manual 
 
CON 11/110 
 
19
 
4.7.3  Calibrating for TDS using TDS standards 
After setting the correct TDS Factor, you can commence calibration in the TDS 
mode. 
From the measurement mode, 
1.  If necessary, press the MODE key to select the TDS mode. 
2.  For the rest of the calibration process, repeat the similar step 2 to 6 as in the 
section of Manual Calibration on page 17, this time using the TDS calibration 
standards. 
NOTE: You can offset the TDS reading up to ± 40% from the default setting. If your 
measured value differs by more than ± 40%, clean or replace electrode as needed.