Instruction Manual 
 
CON 11/110 
 
21
5.2 
Manual Temperature Compensation 
To select the Manual Temperature Compensation (MTC) option, set up the Manual 
Temperature Compensation mode in the SETUP menu Program P 1.3 by setting to 
‘NO”. 
1.  Press SETUP key to enter Set Up mode. The display shows “COF” at “P 
1.0”. 
2.  Press the HOLD/ENTER key to scroll through subgroups until you view 
“ATC” at “P1.3” with small annunciators of “°C” and “ATC” or “°F” and “ATC”. 
3.  Press the HOLD/ENTER key again. The upper display shows “ATC” and the 
lower display shows “YES”.  
4. Press 
the 
MI/▲ or MR/▼ key to select the lower display to show “NO”.  
5.  Press the HOLD/ENTER key to confirm selection and to return to the 
subgroup menu. 
6.  Press the CAL/MEAS key to return to measurement mode.  
 CAL
MEAS
SETUP
YES
°C
ATC
SETUP
°C
HOLD
ENTER
MI
MR
HOLD
ENTER
SETUP
cof
p 1.0
No ‘ATC’
SETUP
READY
ON
p 1.1
SETUP
C
p 1.2
°C
SETUP
A C
p 1.3
°C
 HOLD
ENTER
 HOLD
ENTER
 HOLD
ENTER
SETUP
cof
p 1.0
SETUP
 
Figure 16: Manual Temperature 
Compensation Set Up sequence