Series 100 Plus Option Board Analog Output switch settings
Section 3: Installation
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
3.5.6
Series 100 Plus Option Board Analog Output switch settings
The Series 100 Plus Option Board provides two 4-20 mA analog output signals that are hardware 
configurable to either sink or source current as described in 
 and 
 below 
and in Daniel drawing DE-21056 (
). 
Full HART® functionality is provided so that any commercially available HART® transmitter 
which meets the specifications of the HART® Communications Foundation can be connected to 
the Daniel® Ultrasonic Gas Flow Meter.  
Table 3-42  Series 100 Plus Option board Analog In 2 pressure source setting
Board
Switch
Position
Series 100 Plus Option Board
S13-1
SRC
S13-2
SRC
S13-3
SRC
Table 3-43  Series 100 Plus Option Board Analog Out 1 sink settings
Board
Switch
Position
Series 100 Plus Option Board
S14-1
SINK
S14-2
SINK
S14-3
SINK
If the analog output is configured for ‘source’ mode (i.e., powered by the meter), then it  
must be at the bottom of the loop (stack). The HART® output cannot be multi-dropped 
when configured for source mode.