 Reference, Installation, and Operations Manual
Section 6: Meter operation
3-9000-743 Rev S
June 2013
Volume
 201
Equation 6-19  Base-condition volumetric flow rate
where
6.1.10
Volume
The Daniel Mark III Ultrasonic Gas Flow Meter meter provides forward and reverse volume 
accumulators for each of the three volumetric flow rate values: raw, flow-condition (raw with 
expansion and/or profile correction), and base-condition. Each volume accumulator is actually 
stored as a data pair: (64-bit unsigned integer portion, 32-bit floating point fractional portion). 
For example, a volume of 12345.750 m
3
 is stored as 12345 m
3
 for the integer portion and 
0.750 m
3
 as the fractional portion. Note that while a volumetric flow rate can be positive 
(indicating forward flow) or negative (indicating reverse flow), the volume accumulators are 
always positive values.
The non-volatile volume accumulator data points are as listed in the table below:
=
“raw” volumetric flow rate (m
3
/h) (
QMeter)
=
flow-condition volumetric flow rate (m
3
/h) (
QFlow)
=
base-condition absolute pressure (MPaa) (
PBase)
=
flow-condition absolute pressure (MPaa) 
(
AbsFlowPressure) calculated as shown in 
=
base-condition temperature (K) (
TBase)
=
flow-condition temperature (K) (
FlowTemperature)
=
base-condition compressibility factor (
ZBase)
=
flow-condition compressibility factor (
ZFlow)
Table 6-3  Volume Accumulation Data Points
Volumetric flow 
rate type
Forward direction
Reverse direction
Integer
Fraction
Integer
Fraction
Raw
PosVolUncorr
PosVolUncorrFrac
NegVolUncorr
NegVolUncorrFrac
Flow-condition
PosVolFlow
PosVolFlowFrac
NegVolFlow
NegVolFlowUncorr
Base-condition
PosVolBase
PosVolBaseFrac
NegVolBase
NegVolBaseUncorr
Q
Base
Q
Flow
P
abs f
,
P
abs b
,
---------------
 T
b
T
f
-----
 Z
b
Z
f
-----
=
Q
Base
Q
Flow
P
abs b
,
P
abs f
,
T
b
T
f
Z
b
Z
f