 Reference, Installation, and Operations Manual
Upgrading a Mark II to a Mark III
3-9000-743 Rev S
June 2013
Inactive chords
 309
C.4.8
Inactive chords
The Mark II 
ChordInactv
 (bitfield) parameter is converted to the Mark III 
IsInactiveA...IsInactiveD data points as follows:
C.4.9
AlarmDef
In the Mark II, the 
AlarmDef
 parameter is used for two purposes:
It specifies the number of consecutive batches that a chord must be failed before the 
chord failure is indicated in the 
DataQlty
 register.
It specifies the number of consecutive batches that all active chords must be without 
failure for updating the chord proportions.
The Mark III provides a separate data point for each of these purposes: 
AlarmDef for the first 
purpose and 
PropUpdtBatches for the second. Since the Mark III batch update period is much 
faster than that of the Mark II, the Daniel MeterLink configuration conversion feature does not 
modify the Mark III 
AlarmDef or PropUpdtBatches data points (which should be at default 
values). These points’ values can be read and modified via the Daniel MeterLink 
Edit/Compare 
Configuration
 screen.
Table C-1  Mark II chord inactive conversion to Mark III data points
Mark II ChordInactv
Mark III Data Point
Bit Position
Bit Value
Name
Value
0
0
IsInactiveA
0 (FALSE)
1
1  (TRUE)
1
0
IsInactiveB
0 (FALSE)
1
1  (TRUE)
2
0
IsInactiveC
0 (FALSE)
1
1  (TRUE)
3
0
IsInactiveD
0 (FALSE)
1
1  (TRUE)