 Reference, Installation, and Operations Manual
Upgrading a Mark II to a Mark III
3-9000-743 Rev S
June 2013
Flow pressure and temperature
 311
C.4.11
Flow pressure and temperature
If the Mark II configuration 
EnablePressureInput
 or 
EnableTemperatureInput
 parameter is set to 
0 (None) or 2 (Specified/Fixed), then the Mark II parameter is copied to the corresponding 
Daniel 3410 Series Ultrasonic Gas Flow Meter data point (
EnablePressureInputEnableTem-
peratureInput).
If the Mark II meter is configured for live input (i.e., 
EnablePressureInput
 or 
EnableTemperatureInput
 
parameter is set to 1 (Live)), the conversion depends upon whether or not the Daniel 3410 
Series Ultrasonic Gas Flow Meter supports live input. If the Daniel 3410 Series Ultrasonic Gas 
Flow Meter supports live input (i.e., the Option Board is present and the CPU firmware is 
version 1.30 or later), then the Mark II parameter is copied to the corresponding Daniel 3410 
Series Ultrasonic Gas Flow Meter data point (
EnablePressureInputEnableTemperatureInput). 
Otherwise, the corresponding Daniel 3410 Series Ultrasonic Gas Flow Meter data point is set to 
2 (Fixed).
C.4.12
AGA8 method
The Mark III 
HCH_Method data point allows for the selection of no AGA8 calculations (a 
selection that the Mark II does not provide). 
If the Mark II 
HCH_Method
 parameter is set to External(0) and the Mark II 
SpecRhoMixFlow
SpecZFlow
, and 
SpecZBase
 parameters are equal to zero, then the Mark III 
HCH_Method is set to 
None(4). Otherwise the Mark II 
HCH_Method
 is copied to the Mark III 
HCH_Method data point.
C.4.13
Minimum hold time
For some older Daniel 3410 Series Ultrasonic Gas Flow Meter firmware versions, the hold time 
lower limit is longer than that of the Mark II. Thus, if the Mark II 
MinHoldTime
 parameter is less 
than the Mark III hold time lower limit, the Mark III 
MinHoldTime data point is set to its lower 
limit. Otherwise, the Mark II 
MinHoldTime
 is copied to the Mark III 
MinHoldTime data point.
C.4.14
Log Format parameters
Mark III archive logs (daily, hourly, audit, alarm, and system) are not readable via Modbus. They 
are optionally read using a faster, more efficient method via the Daniel MeterLink 
Logs/Reports - 
Meter Archive Logs
 screen. Se
 for further information on Mark III log retrieval using 
Daniel MeterLink. Thus, the Mark II 
LogDateTimeFormatLogDailyVolumeFormat, and 
LogHourlyVolumeFormat parameters are not converted.