A calibration adjusts the sensor sensitivity to ensure an accu-
rate response to gas.
To calibrate a detector, complete the following:
1.
From the docking module of the applicable detector, 
press C CALIBRATION.
2.
The gas is automatically applied.
3.
When the calibration is complete, the system purges. 
The countdown for time remaining displays beside 
Purge.
 = calibration has passed
 
= calibration has failed
If any test fails, refer to Troubleshooting in the 
MicroDock II Base Station User Manual.
4.
The detector deactivates.
Data Transfer 
(GasAlert Extreme, GasAlertMicroClip, 
GasAlertMax XT only)
This feature transfers datalogs from the detector to the base 
station MultiMediaCard (MMC).
To transfer a datalog, complete the following:
1.
Activate the detector and insert it into the docking 
module. 
2.
Press C DATA TRANSFER. The RUN LED flashes 
rapidly. The base station LCD displays the following 
screen:
Depending upon how many logs are stored on the 
MMC, the data transfer may require 12-14 minutes to 
complete (1-2 minutes for the GasAlertMicroClip and 
GasAlertMax XT).
Note
A maximum of ten datalogs from all modules 
combined can be stored on an MMC (100 datalogs 
for GasAlertMicroClip and GasAlertMax XT).