Honeywell MicroDock II 用户手册

下载
页码 19
1
MicroDock II
Introduction
a
 Warning
To ensure personal safety, read 
 before using the MicroDock II base station.
The MicroDock II Automatic Test and Calibration Station (“the 
base station”) provides automated calibration and bump testing 
for the GasAlert Extreme, GasAlertClip Extreme, GasAlertMi-
cro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, and GasAlert-
Max XT detectors. The system is expandable to include up to 
10 docking modules (maximum six charging docking modules 
plus four non-charging docking modules.
Safety Information - Read First
Use the station only as specified in this guide. Read the follow-
ing Cautions before using the station.
a
 Cautions
• If the base station is damaged or parts are missing, 
contact 
 immediately.
• This equipment uses potentially harmful gas for 
calibration. Use in a well-ventilated area only.
a
 Cautions
• The base station must be attached to a venting system or 
the base station must be used in a well-ventilated area.
• Do not immerse the station in liquids.
• The maximum recommended exhaust line length is 
15.24 m (50 ft.).
• Ensure that the inlet filter is clean.
• Ensure that all gas cylinders contain enough gas.
• A demand flow regulator must be used with all gas 
cylinder connections.
• Calibrate and bump test only in a safe area that is free 
of hazardous gas.
• Do not expose the station to electrical shock or severe 
continuous mechanical shock.
• The base station warranty will be void if the unit is 
disassembled, adjusted, or serviced by non-BW 
Technologies by Honeywell personnel.
• Ensure the exhaust line is not connected to a negative 
pressure system.