Honeywell MicroDock II 用户手册
3
MicroDock II
User Options Menu
User Options Menu
Description
Display
Time/Date modifies the time and
date of the base station.
date of the base station.
Inlet Setup enters the gas type,
the gas concentration levels, and
the gas cylinder lot numbers.
the gas concentration levels, and
the gas cylinder lot numbers.
Pump Setup modifies the station
pump speed. Recommended
pump speeds are 40-45% (350
ml/min.). The pump speed must
be set for each new docking mod-
ule that is added to the station.
NOTE: A flow meter is required.
pump speed. Recommended
pump speeds are 40-45% (350
ml/min.). The pump speed must
be set for each new docking mod-
ule that is added to the station.
NOTE: A flow meter is required.
Contrast brightens or dims the
text of the LCD.
text of the LCD.
Backlight
enables/disables the
LCD backlight. When enabled,
the
the
auto
option automatically
deactivates the backlight when
the base station is not in use.
the base station is not in use.
Description
Display
About
displays the firmware
revision for the base station and
the docking module(s).
the docking module(s).
Format MMC
formats the Mul-
tiMediaCard (MMC). NOTE:
This feature erases all current
data. Refer to the MicroDock II
User Manual.
This feature erases all current
data. Refer to the MicroDock II
User Manual.
Inlet Sel
selects a gas inlet. If
auto
is displayed, the base sta-
tion automatically selects the
correct inlet for the test.
correct inlet for the test.
Pass Code prevents unauthor-
ized access to the menu
ized access to the menu
options. The LCD displays
when it is pass code protected.
Language
displays all LCD
text in one of five languages:
Eng (English), Fran (French),
Deut (German), Port (Portu-
guese), and Espa (Spanish).
Eng (English), Fran (French),
Deut (German), Port (Portu-
guese), and Espa (Spanish).