7
MicroDock II
Using the Base Station
• When the detector has been inserted correctly, the RUN 
LEDs light yellow, Unit Inserted displays on the base 
station LCD, and MicroDock displays on the detector 
LCD. The base station LCD displays the docking mod-
ule number, and the type and serial number of the 
detector.
GasAlertMicroClip
Important: If the GasAlertMicroClip is fitted with a calibration 
cap or an auxiliary filter, it must be removed prior to inserting it 
into the docking module. Refer to the GasAlertMicroClip User 
Manual 
and the MicroDock II Base Station User Manual.
• Insert the bottom of the detector (serial number face up) 
at a 30° angle into the detector bay.
Note
The docking module lid only raises upward 30°. Do 
not force the lid beyond its limit.
• Close the lid and press until the release tabs click. 
• When the detector has been inserted correctly, the RUN 
LEDs on the docking module light yellow and Unit 
Inserted
 displays on the base station LCD. The base 
station LCD displays the docking module number, and 
the type and serial number of the detector.
GasAlertMax XT
• Insert the bottom of the detector (serial number face up) 
at a 30° angle into the detector bay.
Note
The docking module lid only raises upward 30°. Do 
not force the lid beyond its limit.
• Close the lid and press until the release tabs click. Push 
the pump connector towards the detector pump. 
Ensure green is visible in the lower indicator in the 
pump connector. If it is not visible, the MicroDock II base 
station will not recognize the detector.
Note
The detector will briefly alarm when the pump 
connector is indicated. This is normal. The alarm will 
deactivate when the detector is properly inserted into 
the docking module.
• When the detector has been inserted correctly, the RUN 
LEDs on the docking module light yellow and Unit 
Inserted
 displays on the base station LCD. The base 
station LCD displays the docking module number, and 
the type and serial number of the detector.
Using the Base Station
a
 Warning
To prevent possible personal injury and/or property 
damage, only use the station in a safe area that is free of 
hazardous gas.
Ensure that the station is attached to a venting system or 
used in a well ventilated area.