Chart Environmental Chambers 
 
Description of Safety Features 
Description of Safety Features
 
 
On each unit, there are a number of features, which help to ensure safe operation.  These 
features are: 
 
 
❏ 
Temperature High/Low Limit Control: 
 
A user-programmable limit control is provided as an independent measure of safety 
against thermal runaway (hot or cold), which could occur from, but not limited to, the 
following: 
 
Programming error 
 
Primary controller failure  
 
Thermocouple breakage 
 
Heater SSR failure  
 
Circulation blower failure 
 
The limit control is wired into the control devices to provide a totally independent safety 
system.  A separate thermocouple is provided for the limit control and located inside the 
air plenum.  If this device has reached an alarm condition, the system must be reset by 
pressing the “Chamber Reset” button on the control console. 
 
WARNING!
:  The alarm set point of this device should never be set higher 
than +200
°
C or lower than –80°C as these are the 
maximum design temperature of the system. 
 
It is important to note that this safety controller does not protect from a possible failure of 
the cooling solenoid valve.  It is possible for this valve to fail and allow liquid nitrogen to 
continue flowing into the chamber, resulting in extremely cold temperatures.  As a result, 
Chart recommends changing the solenoid valve on a regular basis as a preventative 
maintenance measure against such occurrences.
 
 