Sensors 
  
 
Globe Thermometer 
 
 
Sensors 
 
Globe Thermometer 
The globe thermometer (left position) gives an indication of the radiant heat 
exposure on an individual due to either direct sunlight or hot objects in the 
environment.  This is accomplished by placing a temperature sensor inside a 
blackened copper sphere and measuring the temperature rise.  The WBGT index 
is based on the response of a 6 inch diameter globe.  The QUESTemp uses a 2 
inch diameter globe for a faster response time.  The temperature of the 2 inch 
globe is correlated to match that of a 6 inch globe. 
 
Waterless Wetbulb & Relative Humidity Sensor  
The relative humidity sensor (middle position) is used to calculate the Waterless 
Wetbulb from a combination of dry bulb temperature, humidity and wind speed 
measurements.  The waterless wetbulb is used to calculate an estimated WBGT 
value.   
   
Dry Bulb Thermometer 
The dry bulb thermometer (right position) measures the ambient air 
temperature.  This measurement is used in the outdoor WBGT calculation when 
a high solar radiant heat load may be present.  The series of white plates 
surrounding the sensor shield it from radiant heat. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 1-8:   Sensors identified  
A. Globe thermometer 
B. Relative humidity sensor 
C. Dry bulb thermometer 
 
Sensors