Measurements  
   
WetBulb Globe Temperature (WBGT) 
 
Measurements 
 
The QUESTemp
 48N data logging, with an event logging mode, heat stress 
monitor directly senses three parameters:  dry bulb temperature (DB), globe 
temperature (G), and relative humidity (RH). 
 
It computes the Wet Bulb (WB), the Wet Bulb Globe Temperature (WBGT), and the 
Stay Times Indices for ACGIH, U.S. Navy PHEL charts, and Flag Conditions for U.S.  
Navy/Marine Corp. Ashore. 
 
WetBulb Globe Temperature (WBGT) 
The WBGT is a type of temperature index which combines the effects of 
temperature, humidity, radiant heat, and air flow and provides a relatively simplistic 
tool to analyze thermal comfort.  The resulting WBGT values can then be compared 
to indices of work-rest regimens (stay times) based upon workloads.   
 
The WBGT calculation is a weighted average of the three temperature sensors using 
the following formulas: 
 
 
 
 
 
 
 
 
Stay Times/Rest Times (Indicies) 
Stay Times Indices represent how long a worker should be able to safely work 
under heat stress conditions.  Select one of four indices for displaying and 
printing from the unit: ACGIH Stay Times, NAVY PHEL’s, or Flag Conditions for 
U.S. Navy/Marine Corp. Ashore.  Refer to Appendix B for more information on 
the indices.  NOTE:  it will appear as “Index” on the measurement screen. 
Remote measurements 
 
The top sensor bar may be removed from the instrument and used through a 
remote cable.   (Please refer to Appendix C for remote cable options.)  Shelter 
the instrument and remote the sensor bar if the measured environment is 
expecting heavy rain or if temperatures are above 60
C. 
 
 
**NOTE:  Navy and Marine calculations are all based on the outdoor computation.   
 
 
WBGT (outdoor)  = 0.7WB + 0.2G + 0.1DB (denoted as “WBGT” on the display) 
 