Electronic sensor check  
   
Performing an electronic sensor check 
14 
Electronic sensor check
 
A verification module, Quest model 053-923, may be used to check the operation of the 
QUESTemp’s wet bulb, dry bulb, and globe.   The purpose is to verify that the electronic 
components are within a specific range with known values and a known source.  The 
temperature tolerances should be within +/-0.5
o
C. 
 
 
 
 
 
 
 
 
 
NOTE:   If the sensors are outside of the tolerances, this indicates the sensor alignment 
should be serviced off-site for calibration.  
 
Performing an electronic sensor check 
1.  Ensure the instrument is reading in Celsius prior to your electronic sensor check. 
 
 
  
To change the temperature setting, select Setup from the main menu.  Then, 
either Fahrenheit or Celsius will appear on the screen.  When selected, press I/O 
enter key to switch between settings.  For more information, please refer to page 9, 
step 3.   
 
 
2.  Remove the top sensor bar, place to the side, and plug in the verification module into 
the center pins of the sensor housing.   
 
3.  Verify the measurement readings on the screen are within +/- 0.5
C tolerance to the 
readings printed on the verification module label.  Example below: 
a. Wet Bulb (WB):  11.1
C           b.  Dry Bulb (DB):  45.3C 
c. Globe (G):  69.2
C                   


NOTE: Relative humidity (RH) is not valid on this instrument.  
 
4.  Once completed, remove verification module and place sensor bar back on the 
instrument.  (Tighten down the two bolts.)  Change the Celsius reading back to 
Fahrenheit.  (Refer to step 1 a-b.) 
Electronic Sensor 
check  
 
Example of Verification module