Sensor Alignment  
  
 
RH Sensor Alignment 
15 
 
 Sensor Alignment 
 
For highly accurate measurement readings, you should align your QUESTemp
º 
48N prior to 
data logging or event logging/recording.  For the RH sensor, it is recommended to align with 
a High concentration level using  the 75% sensor salt, NaCl (sodium chloride),  or a Low 
concentration level using the, the 33% sensor salt, MgCl (magnesium chloride). (To order 
sensor alignment salts, refer to an online vendor such as, colepalmer.com.) 
 
RH Sensor Alignment 
1.  To open, navigate to the main menu and select View by pressing I/O Enter key (see 
A).  A measurement screen will appear (see B). 
 
 
 
 
 
 
                   
Figure 1-24:  Main menu with view selected (A) & measurement screen (B) 
 
2.  Press and hold I/O Enter key and then press Down Arrow key from the View menu.  
The Alignment screen will appear. 
 
 
 
 
 
     Figure 1-25:  Alignment screen 
3. 
Press Up/Down Arrow key to select either High or Low.  Then press I/O Enter key.  
 
4. 
Remove (or slide) the Sunshield (white globe) from the RH sensor and place it to the 
side.  (See Figure 1-26). 
5. 
Place a latex finger cot (user supplied) over the humidity sensor in order for the sensor 
to stabilize quicker about 15 minutes (or up to 60 minutes without).  Note:  latex finger cots 
can be found at various hardware stores or online such as colepalmer.com.
  
 
 
 
 
 
 
 
 
 
 
 
 
 
       
Figure 1-26:  Sunshield & placement of finger cot prior to RH alignment 
 
 
“High” is selected when the arrow 
appears in front of the text. 
(RH Sensor) Alignment menu 
Place unrolled finger cot, with a small 
hole cut in the top, so it is completely 
stretched over the humidity sensor. 
 
Sensor alignment  
 
(A) 
(B) 
Note:  WBGT is not used in sensor alignment 
Sunshield:   
Slide off RH 
sensor to 
remove.