6.1.4 Probe
RadioC
Radio probes may only be used in countries in which they have been Type
Approved (see application information of the radio probe).
A radio module (accessory part) is required for the use of radio probes. The
instrument can establish a connection with a maximum of three radio probes. 
Each radio probe has a probe ID (RF ID). This consists of the last 3 digits of the
serial no. and the position of the slide switch (H or L) in the radio probe. 
²
Setting uup aa rradio pprobe: 

A radio module (accessory part) is inserted in the instrument.
º
See chapter 
C
OMMISSIONING
, p. 37.

The configuration menu is open, config. is displayed.

The radio probe is switched on and the transfer rate is set to 2 readings
per second (see the advice on using the radio probe).
1
Probe ¬
¬
RadioC ¬
2
Press 
to select the desired channel no. for the radio probe (P.1,
P.2
or P.3) and confirm with 
- The instrument searches for switched-on radio probes in the receiving
range.
- The probe IDs of the radio probes found are displayed.
If no radio probes were found, this may be because of the following: 
· The radio probe is not switched on or the battery of the radio probe is
spent.
· The radio probe is outside the range of the measuring instrument.
· Sources of interference are influencing the radio transmission (e.g.
reinforced concrete, metal objects, walls or other barriers between
transmitter and receiver, other transmitters of the same frequency,
strong electromagnetic fields).
i
If necessary, rectify the possible causes of the fault in radio
transmission. 
Alternatively, the probe ID can also be entered manually.
i
¬
Press 
to enter the probe ID.
MAN
OK
OK
OK
6. Setting the instrument 43