Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 数据表

产品代码
GSP-830
下载
页码 181
  
SYSTEM 
   133
 
 
2.  Press F1 (Year) to F3 (Second). 
Enter the value using the 
Numerical keys and Enter key. 
BK  SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
   Sec
μ
 
 
The clock display changes 
accordingly. 
Hour
F   1
Minute
F   2
Second
F   3
 
 Hour 
0 to 23 
 Minute 
0 to 59 
 Second 
0 to 59 
Synchronize GSP-830 with Other Devices 
Using the REF INPUT/OUTPUT on the rear panel, the GSP-830 can 
synchronize its internal frequency with other devices. The GSP-830 can 
become the master (reference signal for another device) or the slave (reference 
signal input from another device). 
GSP-830 as master (internal reference signal) 
 
Panel operation 
Connect the reference signal output 
terminal on the rear panel to the 
other device’s reference input. 
 
Signal type 
Output level 
10MHz, 5V TTL signal (assumes 
the load impedance is infinite) 
 Output 
impedance 
50Ω 
Stability check 
1.  Press the System key. 
System
 
 
2.  Press F6 (More). 
More...
F   6
 
 
3.  Press F4 (System Config On). 
System Config
On    Off
F   4
 
 
The system configuration window appears. The Internal 
Reference Signal sign 
  shows the status. 
: enabled,  : disabled. 
 
 
Internal reference signal, standard stability