Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 数据表

产品代码
GSP-830
下载
页码 181
 
 
GSP-830 User Manual 
134
 
 
 
 
Internal reference signal, median stability. 
Available when the ±1ppm stability module 
(page174) is installed. The MED icon 
appears at the bottom of  the display. 
 
 
Internal reference signal, high stability 
(reserved for future use). 
GSP-830 as slave (external reference signal) 
 
Panel operation 
1.  Press the Option key. 
Option
 
 
2.  Press F4 (Ext Ref  Freq) to 
enable external reference signal.
Ext Ref Freq
10MHz
F   4
 
 
3.  Use the Left/Right keys or 
Scroll knob to select the 
external reference frequency. 
 
 
 
4.  Connect the external reference 
signal to the input terminal on 
the rear panel. 
 
 
5.  The external reference signal 
icon appears at the bottom of  
the display. 
 
Signal type 
TTL signal 
Frequency 
10 preset frequencies are available (units in MHz). 
 
1.0 1.544 
2.048 
5.0 10.0 
 
10.24 13.0  15.36 15.4  19.2 
Status check 
1.  Press the System key. 
System
 
 
2.  Press F6 (More). 
More...
F   6
 
 
3.  Press F4 (System Config On). 
System Config
On    Off
F   4
 
 
The system configuration window appears. The 
External Reference Signal sign 
 shows the 
status:   - enabled,   - disabled.