Teledyne API Model 200AH NO
X
 Analyzer Instruction Manual,  06492, Rev. G2 
 
7-14 
Table 7-12:  Converter Efficiency - Automatic Calculation 
Step 
Number 
Action 
Comment 
1. 
Preparation 
Make sure a calibration is completed before setting up 
Converter efficiency. 
2. Press 
 
CAL-CONC-CON-NO2 
Enter the expected NO
2
 concentration, then press ENTR. 
3. 
Input a known 
concentration of NO
2
 
calibration gas. 
Allow reading to stabilize for 10 minutes. 
4. 
Press CAL-ENTR 
By pressing ENTR the M200AH will calculate the ratio of 
the observed to expected NO
2
 and will use that ratio to 
compensate future NO
2
 readings. The calculated value can be 
seen by pressing the SET button. 
NOTE:  The CAL button will come on only if the Instrument is 
measuring NO
2
 gas and ratio is between the limits of 0.80 - 1.20. 
5. 
Press EXIT 
EXIT will return instrument to SAMPLE mode. 
 
Method 2. If the converter efficiency is determined by some external means, the value can be 
directly entered. Press CAL-CONC-CON-SET, then key in the ratio and press ENTR, then 
EXIT. 
7.12. 
  Recommendations for CEM Applications 
7.12.1.   Calibration Gasses 
The following is our recommendation for calibration gasses and equipment for the M200AH. 
Regulator Specifications:  
Ultra high purity, two-stage stainless steel body with stainless steel diaphragms, Tefzel and 
Teflon seats and seals, CGA 660 type. Gauges are 0-4000 psig inlet with 0-100 psig outlet. 
Nitric Oxide Calibration Gas specifications: 
Use NO in nitrogen, EPA protocol 1, NIST traceable (long lead item, order early) for CEMS 
Certification Test gas and acid rain applications.  
Use NO in nitrogen at 80% of calibration range 
 2% for routine calibration. 
Example: Order 80 ppm NO in N2 if calibrating the 0-100 ppm range.